DR 5000 UV-Vis Spectrophotometer |
Product #: DR5000-03 |
**This product was previously sold under Hach part numbers DR5000-01 and DR5000-02. Instrument kit now includes both 115 and 230V power cords.
The DR 5000 UV-Vis Spectrophotometer offers broad spectrophotometric capabilities over the full UV-Vis spectrum:
- - High Speed Wavelength Scanning
- - Touch Screen Display with intuitive user interface
- - Universal sample cell holder holds 5 common cell types
- - USB ports for easy program updates and printer/accessory connections
- - Large sample cell compartment
- - Automatic Test Recognition (when used with TNTplus barcoded vial tests)
Includes: Power cord (115V and 230V), Instrument Manual, Procedure manual (CD-ROM), 1-inch matched glass sample cells,
1-cm matched quartz sample cells and Multi-cell holder
Optional :Accessories for the DR 5000 UV-Vis Spectrophotometer:
- - Sipper Module and Pour-Thru Cell for high volume analysis and trace level measurement
- - Integrated carousel (holds up to seven 1-cm rectangular cells)
DR 5000 UV-Vis Spectrophotometer |
|
Specifications |
|
DR 5000 UV-Vis Spectrophotometer Performance Specifications |
Optical System |
Split Beam |
Operating Mode |
Transmittance (%), Absorbance and Concentration |
Source Lamp |
Tungsten (visible) |
and Deuterium (UV) |
Wavelength Range |
190 - 1100 nm |
Wavelength Accuracy |
± 1 nm |
Wavelength Resolution |
0.1 nm |
Wavelength Calibration |
Automatic |
Wavelength Selection |
Automatic, based on method selection |
Scanning Speed |
One complete scan per minute in 1 nm steps |
Spectral Bandwidth |
2 nm |
Photometric Range |
± 3.0 A in the wavelength range 200 - 900 nm |
Photometric Accuracy |
5 mA at 0.0 - 0.5 A |
1% at 0.50 - 2.0 A |
Photometric Linearity |
< 0.5% at 2 A |
< = 1% at > 2 A |
Stray Light |
KJ-solution at 220 nm |
> 3.3 Abs |
Physical and Environmental |
Width |
17.7 in (450 mm) |
Height |
7.9 in (200 mm) |
Depth |
19.7 in (500 mm) |
Weight |
34.2 Ibs (15.5 kg) |
Operating Requirements |
10 - 40 °C (50 - 104 °F), max. 90% relative humidity (non-condensing) |
Storage Requirements |
-25 - 60 °C (-13 - 140 °F) max. 80% relative humidity (non-condensing) |
Power Specifications |
Mains connection |
100 - 120 V~; 200 - 240 V~; 50/60 Hz; automatic changeover |
Interfaces |
USB 1.1 |
Enclosure rating |
IP 32 |
|
|
|
|